Biasing to increase the ∫idt value of the monitor
The presence of the high permeability magnetic
core and the finite values of load and wire resistance
limit the current-time product (∫idt) a
monitor can measure before its output abruptly
drops towards zero. This drop occurs when the
core material saturates and its permeability approaches
the permeability of air. High current-time products
can be achieved by increasing the core cross section
or by selecting a model with lower sensitivity.
For unipolar pulses, another method exists in biasing. With biasing, a separate
current-carrying conductor is inserted into the monitor opening, as shown in the
figure below, and a DC current flowing in the opposite direction of the primary
current is applied.
Depending on the core material and the selected
model, an increase in the current-time product,
∫idt, on bias current is shown in the graph
below for IPC's standard L models. This graph
shows that for a relatively modest DC current
of 200 mA, the current-time product can be increased
by more than a factor of 2.

For monitors with larger core diameter, such as the H, B or C models, correspondingly
larger currents must be applied to achieve the same improvement.
When applying a bias current it is important that the impedance of the biasing
circuit is high enough not to interfere with the measurement of the primary pulse.
To achieve this a resistance that is at least as high as the table below indicates
must be inserted in the biasing circuit. The biasing current can be increased
by looping several biasing turns through the monitor. In this case, the minimum
resistance values shown in the table must be multiplied by the square of the number
of the bias turns.
| |
Suggested minimum bias resistor for one bias loop.
RTable in Ohms. |
Models with sensitivity of 0.001 V/A |
0.008 |
Models with sensitivity of 0.01 V/A |
0.02 |
Models with sensitivity of 0.1 V/A |
0.066 |
Models with sensitivity of 1 V/A |
7 |
Models with sensitivity of 5 V/A |
160 |
R min = N² bias * R Table
For example, for our CM-10-L, which has a sensitivity
of 0.1 V/A, a minimum resistance of 66 milliohm
must be present in the biasing circuit if one
primary turn is used. On the other hand, if 10
primary turns are used, the minimum resistance
must be 100 times as high, i.e. 6.6Ω for
the above example.
With the minimum resistance in place, the monitor will read 0.5% low. Doubling
Rmin would reduce the error by a factor of two, and the
monitor would read only 0.25% low.
Another type of biasing may be used by applying the bias current to the secondary of the current transformer. When using this method it is important that the impedance of the bias circuit is much higher than the load impedance of the current transformer. IPC would be happy to assist in specific applications.
Return to Current Monitors main page

|